Product overview | Reference users | Brochure & Application Notes

NR01

4 component net-radiation sensor

NR01 is a 4-component net-radiation sensor that is used for scientific-grade energy balance studies. The instrument has separate measurements of solar (Short Wave or SW) and Far Infra-Red (Long Wave or LW) radiation. Major improvements relative to comparable instruments include weight (reduced), solar offsets in the LW signal (reduced), ease of leveling (high, because levelling assembly is included).

Free: For a nice clear print of the original product brochure (PDF) click here.

Figure 1 NR01 4-component net radiation sensor. SW solar radiation sensor or pyranometer (1), LW Far Infra-Red radiation sensor or pyrgeometer (2), radiation shield (3), levelling assembly for x- and y axis (4, 5 and 8).

Figure 2 NR01 top view. Standard cable length is 5 m.  Cable can be installed / replaced by the user. Attachement to a 1 inch tube is extremely easy, levelling possibilities are included. Dimensions in mm.

INTRODUCTION

NR01 serves to measure the 4 separate components of the surface radiation balance. Working completely passive, using a thermopile sensors,  NR01 generates 4 small output voltage proportional to the incoming and outgoing SW and LW fluxes. The SW solar radiation sensors are also called pyranometers, the LW sensors are also called pyrgeometers. For calculation of sky- and surface temperature, a Pt100 temperature sensor is included in the pyrgeometers.

In order to avoid deposition of dew, the pyrgeometers may be heated. A 2-axis levelling assembly is included.

The NR01 cable can easily be installed or replaced by the user. See also RA01 radiameter on the next page, which is a single side version of NR01. Combined with estimates of SW albedo- and of surface temperature, this instrument can also be used for estimation of net-radiation. For a single pyrgeometer see also IR02.

NR01 SPECIFICATIONS

General:

  • Temperature range:                  -40 to +80 oC

  • Range :                                    0 to 2000 Wm-2

  • Temperature sensor:                 Pt100

  • Temperature sensor:                 users own preference can be plugged in

  • Pyranometer (SW):

  • Pyranometer ISO classification:  second class

  • Spectral range:                         305 to 2800 nm

  • Calibration traceability:              WRR

  •  

    Pyrgeometer (LW):

  • Spectral range: 4500 to 50000 nm

  • Calibration traceability: NIST

  • Window heating offset @ 1000 Wm-2 solar radiation: <15 Wm-2
  • heating power: 5 Watt @12VDC

  • OPTIONS

    Additional cable length x metres (add to 5m)

     

     

     

     

     

     


    top of page

     

     