CRC SERIES
CALIBRATION REFERENCE CYLINDERS FOR
THERMAL NEEDLE PROBES TP02/TP08, TP07 and TP09
CRC’s contain traceable thermal
conductivity reference materials for calibration of Non-Steady-State Probes.
There are several different CRC types. The geometry of the different types is
adapted to fit needles with a certain diameter.
Free: For a nice clear print of the original product
brochure (PDF) click here.

Figure 1: generalised picture of
CRC Calibration Reference Cylinder. Explanation can be found in the text.
INTRODUCTION
- The hart of the CRC cylinders is a
specially manufactured amorphous plastic reference material (5). Contrary to
other plastics, this material does not have any directionally dependent
thermal properties.
- In the plastic material a hole is
drilled (4), with the approximate outer dimensions (ND Needle Diameter and IL
Insertion Length) of the needle that needs to be calibrated.
- The plastic material is incorporated in
a metal protection casing (2). A plug (7) is provided for sealing purposes.
- In order to guarantee perfect contact
between needle (1) and reference material (5), the metal casing is filled
with a fluid, glycerol (3).
- An additional hole (6) is provided in
order to be able to perform checks of material stability with thin needle
probes.
- The user is recommended to send the CRC
to Hukseflux for a check of material stability every 5 years.
-
- Hukseflux is a leading supplier of
Non-Steady-State Probes, and is the inventor of the CRC calibration method.
-
- NOTE: As a low-cost (lower accuracy)
alternative reference material Hukseflux promotes use of glycerol filled with
plastic fibre material. This is delivered as a standard with most Hukseflux
systems.
- Type
|
- Dimensions
|
- Suitability for needle types
|
- CRC01
|
- H = 260
- D = 100
- IL = 185 - 30
- ND = 1.6
|
- TP02, TP08
|
- CRC04
|
- H = 206
- D = 100
- IL = 137 - 6
- ND = 3.6
|
- TP07
|
- CRC05
|
- H = 260
- D = 100
- IL = 190 - 12
- ND = 6.6
|
- TP09
|
Table 1: CRC cylinder types and
their application
CRC SPECIFICATIONS
-
Test method: ASTM D
5334-00 and D 5930-97, IEEE Std 442-1981
-
Temperature range: 0 to 60 degrees C
-
Thermal conductivity
of reference: 0.2 W/mK (nominal value)
-
Traceability: to NPL
National Physical Laboratory, United Kingdom
|