TP02 is very often used with Campbell Scientific CR10X. A manual
on application of TP02 with CR10X is obtainable here. More details on
the science: non-steady state probe science. For
a free copy of the product brochure click here
for the PDF version. For a smaller version of
TP02, see TP08.

Figure 1:
TP02 Non-Steady-State Probe consists of a needle with 2 thermocouple
junctions (3), (4) (the tip acts as a reference) and a heating wire, (2). It is
inserted into the medium that is investigated. In the base, (6), a temperature
sensor, (1), is mounted. Advantages of this design: an optimal accuracy
independent of the medium temperature, a minimal sensitivity to thermal
gradients, a high sensor stability and the possibility to use normal cables and
connectors. All dimensions are in mm.
Introduction
The standards of the American Society
for Testing and Materials (ASTM) and Institute of Electrical and Electronics
Engineers (IEEE) describe proven methods for determining the thermal
conductivity of materials.
ASTM D 5334-00 and D 5930-97 and IEEE
Std 442-1981 " Standard Test Methods" specify the use of Non-Steady-State Probes
(NSSP) in various applications. In general a NSSP consists of a heating wire,
representing a perfect line source, and a temperature sensor capable of
measuring the temperature at this source. The probe is inserted in the medium
that is investigated. The NSSP principle relies on a unique property of a line
source: after a short transient period the temperature rise, ∆T, only depends on
heater power, Q, and medium thermal conductivity, λ:
∆T = (Q / 4 π λ) (ln t + B)
With ∆T in K, Q in W/m, λ in W/mK, t the
time in s and B a constant. By measuring the heater power, and tracing the
temperature in time (for TP02 typically during 2 minutes), λ can be calculated.
The sample size is not critical, as long as a 20 mm radius around the needle is
covered.
The measurements of Q, t and ∆T are all
direct measurements of power, time, and temperature respectively. These are done
without need of reference materials. The measurement with TP02 is absolute.
TP02 can be obtained as a separate probe
for incorporation in the user's measurement and control system. It is suitable
for operation with the Campbell Scientific CR10X.
TP02 Design
The Hukseflux TP02 is designed to be
able to cover a wide range of applications. Design considerations are as
follows:
- Optimal Accuracy:
TP02 incorporates 2 thermocouple junctions in the needle producing a voltage
output, U, that is proportional to ∆T. The one in the tip is not heated, and
serves as a reference junction.
- In this configuration the voltage before
starting the measurement is always small, independent of the medium
temperature. In designs with only one sensor in the needle, the signal U will
appear on top of a larger signal.
- The two junction design has a superior
accuracy when measuring at high and low temperatures.
- In addition, by having both the
reference- and the hot junction in the needle, the sensitivity to temperature
changes of the medium is minimised.
Temperature sensor in the base: The sensor in
the base (Pt1000) serves as a cold junction measurement for establishing the
absolute medium temperature T. (As required by ASTM). This is used for
temperature correction of the thermocouple sensitivity, again offering superior
accuracy across the temperature range.
Temperature and moisture resistance: All the
materials in TP02 are highly temperature resistant. It has a full stainless
steel needle and base, with a welded tip (IP68) and a potted cable (IP67). The
needle is 100% waterproof.
Standard
cables and connectors: Putting the Pt1000
temperature sensor in the base allows the use of normal copper core cables and
connectors for cable extension of TP02.
Long term
stability: The fully sealed construction is
essential for prevention of internal corrosion and maintaining stability of the
sensor over time.
Data
processing: the data obtained with TP02 can be
processed in any normal spreadsheet program.

Figure 2
The signal of TP02 as a function of the natural logarithm of time (ln(t)).
After a transient period the graphs show linear behaviour. In this phase the
slope of the graph is inversely proportional to the thermal conductivity λ.
Calibration/ ISO 9000
Verification of the stability of the
total probe can be done by repeated (yearly) testing in glycerol, preferably at
several temperatures. Alternatively Calibration Reference Cylinders (CRC)
traceable to NPL can be purchased at Hukseflux. TP02 is suitable for use by ISO
certified labs.
Suggested use
More information/ Options
Standards: ASTM standards can be
obtained from ASTM at http://www.astm.org.
Alternative designs: Hukseflux is
specialised in NSSP design. Alternative models, for instance smaller, more robust or
temperature resistant have been manufactured on request.
Permanent installation in soils: The
dedicated model TP01 is specifically designed for long term monitoring. A
separate brochure is available.
Solutions for measurement and control:
See the TP02 manual or inquire at Hukseflux. A program for Campbell Scientific
CR10X is available. Turn key systems are offered for sale: see TPSYS.
Manuals: The
TP02 manual is available free of charge as a PDF file via
e-mail.
NEW!
- For high accuracy calibration CRC
Calibration Reference Cylinders are available.
- For insertion into hard soils
GT Series Guiding tubes can be applied.
TP02 Specifications
-
Test method: ASTM D
5334-00 and D 5930-97 IEEE Std 442-1981
-
Needle length: 150 mm
-
Traceability: NPL
-
Range (λ): 0.1
to 6 W/m.K
-
Sensitivity (∆T): Thermocouple
K, ANSI MC96.1-1982
-
Temperature range (total
sensor -55 to +180 °C including cable):
-
Accuracy (@ 20 °C): +/- (3% + 0.02) W/mK
-
Temperature dependence
of the accuracy: +/- 0.02 %/K (additional)
-
Measurement cycle duration: 200 s (typical)
-
Power requirements (switched) : 3 V, 1 Watt
(max)
-
Medium / sample
requirements: Granular materials, powders, slurries, gels, pastes. Some
samples require pre-drilling. Sample size: Min 20 mm radius. Smaller samples:
consult HUkseflux. Caution: please
consult the product manual for more details.
-
Protection needle and base: IP 68
-
Protection total sensor: IP 67